SE-2003&SE-2012 Series Holter System Recorder User Manual
- 7 -
Chapter 2 Introduction
The manual mainly describes how to operate and maintain SE-2003/SE-2012 series Holter
System Recorder (thereafter referred as SE-2003/SE-2012 series). The SE-2003/SE-2012 series
Holter Analysis System consists of the recorder and analysis software. For the operation of
analysis software, refer to software manual.
The SE-2003/SE-2012 series have four recorder models: SE-2003, SE-2012, SE-2003A, and
SE-2012A. The difference lies in the SD card. SE-2003 and SE-2012 recorders can be inserted
with micro-SD card while SE-2003A and SE-2012A inserted with SD card.
Ambulatory Electrocardiogram (Holter) detection technology is an effective tool to detect
cardiovascular without the inference of distance, time, environment, the restriction of body
position and activity. It can detect large amount of ECG information and is unique in capture of
transient myocardial ischemia and diagnosis of transient arrhythmia.
SE-2003/SE-2012 series is powered by an AAA battery. SD card is used as storage media and
LCD screen is used to set parameters and check wave quality. With a normal AAA alkaline
battery, SE-2012/SE-2012A can continuously record non-compressed and full-disclosure 12-lead
ECG data for 2448 hours or 3-channel ECG data for 24~96 hours . With a normal AAA
alkaline battery, SE-2003/SE-2003A can continuously record non-compressed and full-disclosure
3-channel ECG data for 24~96 hours. With a Li-Fe battery, SE-2003/SE-2003A can continuously
record non-compressed and full-disclosure 3-channel ECG data for 7 days. SE-2003/SE-2012
series can detect and record pacemaker pulse information.
The Holter System is intended to be used in hospitals and clinics by trained personnel under the
direction of doctors. The patient wears the Holter Recorder with the help of the doctor and the
recorder records ECG data for at least 24 hours continuously. The patient may return home
during the recording process. Finally, the doctor analyzes the ECG data recorded and makes
diagnosis.
NOTE:
1. Long-time recording analysis should be supported by software with corresponding
functions.
2. The manufacturer recommends the type of LiFe battery that can be used.
Recommended LiFe battery: ENERGIZER L92
3. The manufacturer does not provide Ni-MH rechargeable battery and its charger.