SE-2003&SE-2012 Series Holter System Recorder User Manual
- 22 -
Chapter 5 System Setting
After setting the parameters, press Exit to return to upper level menu.
NOTE: The underlined options are the default value.
5.1 Basic Setting
For SE-2003 or SE-2012 connected with 5- or 7-electrode patient cable, the
number of recording channels can be set to 3, 2 or 1.
Choose from: 0.5, 1, or 2.
The default startup setting is the value adjusted last time.
Set the duration for the data sampling process.
For SE-2003/SE-2003A, choose from: 1 day, 2 days, 3 days, 4 days or 0.
For SE-2012/SE-2012A, choose from: 1 day, 2 days or 0.
0 means no recording time limit. The recording won't stop until the battery
runs out.
ECG preview can be set to On or Off.
If set to Off, ECG preview will be skipped.
Patient ID can be set to On or Off.
When set to On, you need to input the patient ID before recording.
NOTE: If the patient has been registered in the analysis software
beforehand, the recorder enters the recording screen directly.
ECG display during recording process can be set to On or Off.
Set the value as On and enter ECG display mode by pressing Left-up
key and confirmation key during recording process and observe
waveform.
It is unable to enter ECG display mode by pressing Left-up key and
confirmation key during recording process if the value is set to Off. This
helps to prevent unnecessary influence to patients.
Whether entering ECG display mode or not, recording signals will not be
influenced.
Data protection is used to detect if the recorded data has been analyzed.
The default value is On.
If it is set to On, the recorded data has to be analyzed by analysis software
at least once before starting the next recording.