SE-2003&SE-2012 Series Holter System Recorder User Manual
- 26 -
The validated cleaning agents for cleaning the device and reusable accessories are:
Mild near neutral detergent
Ethanol (75%)
Isopropanol (70%)
Cleaning agents should be applied or removed using a clean, soft, non-abrasive cloth or paper
towel.
7.2.1 Cleaning the Recorder
WARNING
Turn off the power and take out the battery before cleaning.
1. Switch off the recorder and take out the battery.
2. Wipe the exterior surface of the equipment using a soft cloth dampened with the cleaning
solution until no visible contaminants remain.
3. Wipe off the cleaning solution with a fresh cloth or towel dampened with tap water after
cleaning until no visible cleaning agent remains.
4. Dry the recorder in a ventilated and cool place.
7.2.2 Cleaning the Patient Cable
1. Wipe the patient cable with a soft cloth dampened with the cleaning solution until no visible
contaminants remain.
2. Wipe off the cleaning solution with a fresh cloth or towel dampened with tap water after
cleaning until no visible cleaning agent remains.
3. Wipe off with a dry cloth to remove residual moisture.
4. Leave the patient cable to air dry.
CAUTION
Any remainder of cleaning solution should be removed from the recorder and the patient
cable after cleaning.
7.3 Disinfection
To avoid permanent damage to the equipment, it is recommended that disinfection is performed
only when it is considered as necessary according to your hospital' regulations.
Clean the equipment and reusable accessories before they are disinfected. The validated