SE-2003&SE-2012 Series Holter System Recorder User Manual
- 17 -
Chapter 4 ECG Sampling
4.1 Overview
The operation steps of SE-2003/SE-2012 series recorder are as follows:
1. Decide recording plan, such as lead mode, pacemaker pulse data needed or not;
2. Insert SD card into card-reader of analysis system or directly connect it with analysis system
via USB cable to register E-label. The registered SD cards prevent possible confusion caused
by recording of different patients at the same time;
3. Place electrodes;
4. Load a full and fresh AAA battery;
5. Set recording parameters through recorder keyboard and LCD;
6. Preview ECG wave through LCD and check whether electrodes are properly placed and
make adjustment if necessary;
7. Start to record through menu when all steps above are ok;
8. Begin to record formally after recorder initialization;
9. Check wave quality of ECG signal during recording process through LCD and check
placement quality of electrodes;
10. End of recording.
NOTE:
1. SE-2012/SE-2012A supports three kinds of lead mode: 10-electrode standard
12-lead mode, 7- electrode-3-channel mode and 5-electrode-3-channel mode.
Recorder will automatically identify and record data generated from different leads
accordingly.
2. SE-2003/SE-2003A supports 7-electrode or 5-electrode 3-channel mode.

Related product manuals