SE-2003&SE-2012 Series Holter System Recorder User Manual
- 27 -
disinfectants for disinfecting the device and reusable accessories are:
Ethanol (75%)
Isopropanol (70%)
If Ethanol or Isopropanol is used for both cleaning and disinfecting, then a new cloth is required
to be used for the disinfection step.
CAUTION
1. Do not use high-temperature, high-pressure vapour or ionizing radiation as
disinfection methods.
2. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.
3. Clean and disinfect reusable electrodes after each use.
7.3.1 Disinfecting the Recorder
WARNING
Turn off the power and take out the battery before disinfection.
1. Switch off the recorder and take out the battery.
2. Wipe the exterior surface of the equipment using a soft cloth dampened with the disinfectant
solution.
3. Wipe off the disinfectant solution with a dry cloth after disinfection if necessary.
4. Dry the recorder for at least 30 minutes in a ventilated and cool place.
7.3.2 Disinfecting the Patient Cable
1. Wipe the patient cable with a soft cloth dampened with the disinfectant solution.
2. Wipe off the disinfectant solution with a dry cloth after disinfection.
3. Leave the patient cable to air dry for at least 30 minutes.
7.4 Care and Maintenance
7.4.1 Visual Inspection
Perform a visual inspection of all equipment and peripheral devices daily. If you notice any items
that need repair, contact a qualified service engineer to make the repairs.