Acclarix AX8/Acclarix AX7 Diagnostic Ultrasound System User Manual Utilities
- 152 -
11.1.4. Miscellaneous Set-up
Figure 11-4 Miscellaneous Set-up Screen
Item
Options
Description
Default review layout
4*4, 2*2, 1*1, 2*3
Set the default review layout.
Overlay
Show or hide overlay icons on thumbnail images.
PW
mode
Gain
Optimize Gain or not when press Auto button on
the control panel in PW mode.
DR
Optimize Dynamic Range or not when press
Auto button on the control panel in PW mode.
Scale
/Baseline
Optimize Scale/Baseline or not when press Auto
button on the control panel in PW mode.
US Support
Set whether the server in USA or in China will be
connected when using the Remote Support
function. Checking US support will display a " US
Support" button on the touch screen providing
the access to connecting the server in USA.
Otherwise, a "CN Support" button will be
displayed providing the access to connecting the
server in China. Contact EDAN Service for the
instructions for using Remote Support function.
Warning: Remote Support is only a training and
service tool. It is not intended for diagnosis.
Table 11-4 Miscellaneous Set-up Information