Acclarix AX8/Acclarix AX7 Diagnostic Ultrasound System User Manual Imaging
- 37 -
Adjusts the current map for the active color variation.
Persistence determines the number of frames that are averaged
together for display. Levels of Off, Low, Med and High are
available.
The smoothness filter determines the spatial filtering that is applied
to the Color image. Higher filter levels create a smoother image.
Upward presses increase the filter. Downward presses decrease
the filter.
When the system receives both B-mode and color signals from a
region within the Color ROI box, the Threshold determines
whether to display overlapping signals as grayscale or color.
In Color mode, higher Threshold values display more color, and
lower Threshold values display more grayscale.
In TDI mode, higher Threshold values display more grayscale, and
lower Threshold values display more color
Upward presses increase threshold. Downward presses decrease
threshold.
Frequency determines the transmit frequency used by color
Doppler. Upward presses increase the frequency. Downward
presses decrease the frequency.
This control is only available for linear transducers. It steers the
Color ROI box angle left or right.
Color Doppler supports image presets for Low Flow, Medium Flow,
and High Flow.