Acclarix AX8/Acclarix AX7 Diagnostic Ultrasound System User Manual Transducers and Biopsy
- 71 -
instructions provided by the disinfectant manufacturer for the concentration of the disinfection
solution, method of dilution, method of disinfection, temperature and cautions during use.
5. Place the cleaned and dried transducer in contact with the disinfectant (refers to figure 6-3 for the
contacting area) for the time specified by the disinfectant manufacturer. For example, the contact
time recommended by the manufacturer for soaking in Cidex Glutaraldehyde(2.4%) is at least 45
min.
6. Rinse the transducer thoroughly with sterile water to remove all chemical residues. For example,
it is required to flush the transducer after soaking in Cidex Glutaraldehyde(2.4%) with plenty of
sterile water (about 2 gallons) at least one time. Or, follow the complete rinsing instructions
provided by the disinfectant manufacturer to rinse the transducer.
7. Wipe off the water on the transducer with a dry sterile cloth. Leave the transducer to air dry.
8. Inspect the transducer to ensure that there is no damage.
Figure 6-3 Depth of the Transducer Immerged into Disinfectant
WARNING
1. Do not immerse the transducer connector. If the cable connector is immersed, do not plug the
connector into the system. Rinse the connector under running water and dry it thoroughly. If
necessary, contact EDAN for service.
2. Prohibit infiltration of any type of liquid into the device or the transducer.
3. Do not immerse the cable and connector of the transducer into solutions. Transducers can be
submerged to, but not including, the strain relief of the transducer array. Do not immerse or soak
any part of a transducer in any cleaning material not listed in the recommended list of disinfectant.
4. Use the immersion method to disinfect the intra-cavitary and intra-operative transducers.
5. Only non-immersion method can be used with solution of ethanol or isopropanol.
6. The immersion time should not exceed the time that is specified by the disinfectant manufacturer.
L17-7SQ
Disinfectant
E8-4Q
E10-3HQ
E10-3BQ

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN Acclarix AX8 and is the answer not in the manual?

EDAN Acclarix AX8 Specifications

General IconGeneral
BrandEDAN
ModelAcclarix AX8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals