Patient Monitor User Manual Product Specifications
- 259 -
IBP
Measure
Measuring
Range
Art 0 mmHg to + 300 mmHg
PA -6 mmHg to + 120 mmHg
CVP/RAP/LAP/ICP -10 mmHg to + 40 mmHg
P1/P2 -50 mmHg to + 300 mmHg
Resolution 1 mmHg
Accuracy (not including sensor)
±
2% or
±
1 mmHg, whichever is
greater
ICP:
0 mmHg to 40 mmHg:
± 2% or ±1
mmHg, whichever is greater;
-10 mmHg to -1 mmHg: undefined
Pressure Unit kPa, mmHg, cmH
2
O
Pressure sensor
Sensitivity 5 μV/V/mmHg
Impedance Range 300 to 3000
Filter DC~ 12.5 Hz; DC~ 40 Hz
Zero
Range:
±
200 mmHg
Pressure Calibration
Range
IBP (excluding ICP) 80 mmHg to 300 mmHg
ICP 10 mmHg to 40 mmHg
Volume Displacement 7.4 x 10
4
mm
3
/ 100mmHg
A.18 CO
2
Complies with ISO 80601-2-55: 2011.
EDAN Module
Intended
Patient
Adult, pediatric, neonatal
Measure
Parameters
EtCO
2
, FiCO
2
, AwRR
Unit mmHg, %, kPa
Measuring
Range
CO
2
0 mmHg to 150 mmHg (0% to 20%)
AwRR 2 rpm to 150 rpm
Resolution
EtCO
2
1 mmHg
FiCO
2
1 mmHg