Patient Monitor User Manual Product Specifications
- 260 -
AwRR 1 rpm
Accuracy
EtCO
2
± 2 mmHg, 0
mmHg to 40
mmHg
Respiratory
rate 60 rpm
Typical conditions:
Ambient temperature: (25
± 3)
°C
Barometric pressure: (760
±
10) mmHg
Balance gas: N
2
Sample gas flowrate: 100
ml/min
± 5% of reading,
41 mmHg to 70
mmHg
± 8% of reading,
71 mmHg to 100
mmHg
± 10% of reading,
101 mmHg to 150
mmHg
± 12% of reading
or
± 4 mmHg,
whichever is
greater
Respiratory
rate > 60 rpm
All conditions
AwRR
± 1 rpm
Drift of
Measure
Accuracy
Meets the requirements of the measure accuracy
Sample Gas
Flowrate
70 ml/min or 100 ml/min(default), accuracy:
± 15 ml/min
Warm-upTime Display reading within 20 s; reach to the designed accuracy within 2 minutes.
Rise Time
< 400 ms (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 500 ms (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Response
Time
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Work Mode Standby, measure
O
2
Compensation
Range: 0% to 100%
Resolution: 1%
Default: 16%
N
2
O
Compensation
Range: 0% to 100%
Resolution: 1%
Default: 0%
AG
Compensation
Range: 0% to 20%
Resolution: 0.1%
Default: 0%