Patient Monitor User Manual Product Specifications
- 261 -
Humidity
Compensation
Method
ATPD (default), BTPS
Barometric
Pressure
Compensation
Automatic (The change of barometric pressure will not add additional errors
to the measurement values.)
Zero
Calibration
Support
Calibration Support
Alarm
EtCO
2
, FiCO
2
, AwRR
Apnea
Alarm Delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s, 60s; default value is 20 s.
Data Sample
Rate
100 Hz
EtCO
2
Change
1
AwRR 80rpm, meet the accuracy
mentioned above;
AwRR > 80 rpm, EtCO
2
descends 8%;
AwRR > 120 rpm, EtCO
2
descends 10%
with 2 m gas sampling tube,
sample gas flowrate: 100
ml/min)
AwRR 60rpm, meet the accuracy
mentioned above;
AwRR > 60 rpm, EtCO
2
descends 8%;
AwRR > 90 rpm, EtCO
2
descends 10%;
AwRR > 120rpm, EtCO
2
descends 15%;
with 2 m gas sampling tube,
sample gas flowrate: 70
ml/min)
Note 1: Use a test device equivalent to EN ISO 80601-2-55 fig 201.101 to measure at 1:2 I/E
ratio. Respiration rate accuracy is determined by frequency of device, and end-tidal reading
change refers to the nominal value.
Interfering Gas Effects:
Gas Gas Level (%) Quantitative Effect/Comments
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
60
4
5
5
5
15
The interfering gas will have no effect on the
measurement value if compensation of O
2
, N
2
O,
anesthetic agents has been correctly set.
Respironics Module
Applicable Patient Type Adult, pediatric and neonatal patients
Technique Infra-red Absorption Technique