Patient Monitor User Manual Product Specifications
- 261 -
Humidity
Compensation
Method
ATPD (default), BTPS
Barometric
Pressure
Compensation
Automatic (The change of barometric pressure will not add additional errors
to the measurement values.)
Zero
Calibration
Support
Calibration Support
☆Alarm
EtCO
2
, FiCO
2
, AwRR
☆ Apnea
Alarm Delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s, 60s; default value is 20 s.
Data Sample
Rate
100 Hz
EtCO
2
Change
1
AwRR ≤ 80rpm, meet the accuracy
mentioned above;
AwRR > 80 rpm, EtCO
2
descends 8%;
AwRR > 120 rpm, EtCO
2
descends 10%
with 2 m gas sampling tube,
sample gas flowrate: 100
ml/min)
AwRR ≤ 60rpm, meet the accuracy
mentioned above;
AwRR > 60 rpm, EtCO
2
descends 8%;
AwRR > 90 rpm, EtCO
2
descends 10%;
AwRR > 120rpm, EtCO
2
descends 15%;
with 2 m gas sampling tube,
sample gas flowrate: 70
ml/min)
Note 1: Use a test device equivalent to EN ISO 80601-2-55 fig 201.101 to measure at 1:2 I/E
ratio. Respiration rate accuracy is determined by frequency of device, and end-tidal reading
change refers to the nominal value.
Interfering Gas Effects:
Gas Gas Level (%) Quantitative Effect/Comments
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
60
4
5
5
5
15
The interfering gas will have no effect on the
measurement value if compensation of O
2
, N
2
O,
anesthetic agents has been correctly set.
Respironics Module
Applicable Patient Type Adult, pediatric and neonatal patients
Technique Infra-red Absorption Technique