Patient Monitor User Manual Alarm Information
- 58 -
Message Cause Alarm Level Action Taken
NIBP Aux Excessive
Pressure
Pressure has exceeded the
second safety limit as
specified.
High
Notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Time Out
Measuring time has
exceeded the specified time.
Low
Measure again or
use other measuring
method.
NIBP Self Test
Error
Sensor or other hardware
errors.
Low
If failure persists,
stop using
measuring function
of NIBP module
and notify
biomedical engineer
or manufacturer’s
service staff.
NIBP Cuff Type
Error
The cuff type used isn’t
consistent with the patient
type.
Low
Confirm the patient
type and change the
cuff.
NIBP System
Pressure
Abnormality
Atmospheric pressure or
system pressure is
abnormal. The valve is
occluded so that deflation is
failed.
Low
Check whether the
airway is occluded
or pressure sensor
works properly. If
the problem still
exists, contact your
service personnel.
NIBP System
Failure
NIBP is not calibrated. High
Contact your
service personnel.
NIBP Weak Signal
Cuff is too loose or patient
pulse is too weak.
Low
Use other methods
to measure blood
pressure.
NIBP Range
Exceeded
Maybe the patient blood
pressure value is beyond the
measurement range.
High
Use other methods
to measure blood
pressure.
NIBP Loose Cuff
Cuff is not properly
wrapped or no cuff is
connected.
Low
Properly wrap the
cuff.