Patient Monitor User Manual Alarm Information
- 59 -
Message Cause Alarm Level Action Taken
NIBP Interference
Signal noise is too large or
pulse rate is not regular due
to the patient movement.
Low
Make sure that the
patient under
monitoring is
motionless.
NIBP Leak Test
Error
Fail to deflate normally
during the leak test, so
NIBP leak test cannot be
finished.
Low
Test again. If the
problem still exists,
contact your service
personnel.
TEMP
TEMP T1 Sensor Off
Temperature cable of TEMP
channel1 may be
disconnected from the
monitor.
Low
Make sure that the
cable is properly
connected
TEMP T2 Sensor Off
Temperature cable of TEMP
channel2 may be
disconnected from the
monitor.
Low
Make sure that the
cable is properly
connected.
Excessive T1
TEMP1 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
Excessive T2
TEMP2 measuring value is
beyond measuring range.
High
TEMP Comm Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
T1 Calibration
Failed
T1 calibration failed.
High
Please check
whether the module
works properly.
T2 Calibration
Failed
T2 calibration failed
High
IBP
YY Sensor Off (YY
stands for the IBP
label name: Art, PA,
CVP, RAP, LAP, ICP,
P1 and P2)
IBP sensor falls off. Medium
Check the sensor
connection and
reconnect the sensor.