Patient Monitor User Manual Monitoring Quick TEMP
- 138 -
WARNING
8 Make sure disposable TEMP sensor covers are used to limit patient
cross-contamination. The use of any other probe cover may produce temperature
measurement errors or result in inaccurate readings.
9 Quick Temp measurement isn’t suitable for use during defibrillation.
NOTE:
1 After one measurement, the user should put the sensor well to the sensor bracket
and then take it out for starting a new measurement.
2 To ensure optimal accuracy, always confirm that the correct measurement position is
selected.
3 The reference body site temperature is the same as the temperature of the
measuring site.
15.3 Measuring Procedure
15.3.1 Measurement for Oral Temperature
1. Ensure the oral probe (white probe) and probe are well installed.
2. Remove the probe from the probe well.
3. Observe the oral mode indicator on the screen (flashing head icon).
If this icon is not flashing, press the
Measure Pos button and set it to Oral until the head icon
appears.
4. Load the probe cover.
5. Place the probe tip deep into the patient’s sublingual pocket as shown in the following figure.
Measuring Position in Mouth
Sublingual Pocket