Patient Monitor User Manual Product Specification
- 238 -
A Product Specification
NOTE:
The performance of the equipment with mark is determined to be essential
performance.
A.1 Classification
Anti-electroshock type Class I equipment and internal powered equipment
Anti-electroshock degree CF: ECG (RESP), TEMP, IBP, C.O., Quick Temp
BF: SpO
2
, NIBP, CO
2
, AG
Ingress Protection IPX1
Disinfection/sterilization method Refer to Chapter Care and Cleaning for details.
Working system Continuous operation equipment
Compliant with Standards IEC 60601-1: 2005+A1 :2012; IEC 60601-1-2: 2014;
EN 60601-1: 2006+A1 :2013; EN 60601-1-2: 2015;
IEC 60601-2-49: 2011
A.2 Physical Specifications
A.2.1 Size and Weight
Product Size Weight (standard
configuration, without
battery)
iM50/M50 261 mm (W) × 215 mm (H) ×198 mm (D) < 3.6 kg
iM60 303 mm (W) × 254 mm (H) × 161 mm (D) < 3.8 kg
iM70 328 mm (W) × 285 mm (H) × 158 mm (D) < 4.5 kg
iM80/M80 370 mm (W)× 320 mm (H) × 175 mm (D) < 7 kg
A.2.2 Environment Specification
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all
products.
Temperature
Working
+0
°C to +40 °C(32 ~104 )