Patient Monitor User Manual Networked Monitoring
- 40 -
4.7 HL7 Communication
The monitor supports HL7 protocol to upload data. Select Menu > User Maintain and input user
maintain password > Security. In this menu:
Set HL7 to On/Off. To avoid hacker attacking, setting HL7 to Off is normally
recommended.
Set HL7 Encryption method to Off or TLS (default).
To set HL7 IP address of the client-side, select User Maintain > Network Maintain.
Click Import Certificate to install/upgrade the Certificate via USB flash drive. The certificate
issued by Certificate Authority (CA) is recommended and self-signed certificate should be
avoided. For detailed steps of importing certificates, please refers to service manual.
For more information about HL7 communication, refer to HL7 Communication Protocol Service
Manual.