EasyManua.ls Logo

EDAN iM70 - A.5 RESP

EDAN iM70
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual Product Specification
- 250 -
QT/QTC (ms)
P-QRS-T AXIS
A.5 RESP
Method Impedance between RA-LL, RA-LA
Measurement lead Options are lead I and II. The default is lead II.
Calculation Type Manual, Automatic
Baseline Impedance Range 200 to 2500 (with ECG cables of 1 K resistance)
Measuring Sensitivity Within the baseline impedance range: 0.3
Waveform Bandwidth 0.2 Hz to 2.5 Hz (-3 dB)
Respiration Excitation
Waveform
Sinusoid, 45.6 kHz (
±
10%), < 350 μA
RR Measuring Range
Adult
0 rpm to120 rpm
Neo/Ped
0 rpm to150 rpm
Resolution 1 rpm
Accuracy
Adult
6 rpm to 120 rpm:
±
2 rpm
0 rpm to 5 rpm: not specified
Neo/Ped
6 rpm to 150 rpm:
±
2 rpm
0 rpm to 5 rpm: not specified
Gain Selection
×0.25, ×0.5, ×1, ×2, ×3, ×4, ×5
Sweep
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Apnea Alarm Time Setup
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
A.6 NIBP
Complies with IEC 80601-2-30: 2009+A1: 2013.
EDAN Module
Technique Oscillometry

Table of Contents

Related product manuals