Patient Monitor User Manual Product Specification
- 249 -
Accuracy of Heart Rate
Meter and Response to
Irregular Rhythm
Complied with IEC 60601-2-27: 2011, Sect. 201.7.9.2.9.101 b)
4), the HR value after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy: 80 bpm±1 bpm
Slow alternating ventricular bigeminy: 60 bpm±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm±1 bpm
Bidirectional systoles: 91 bpm±1 bpm
Time to Alarm for Heart
Rate alarm conditions
Asystole alarm: 10 s
HR low alarm: 10 s
HR high alarm: 10 s
Arrhythmia analyses
(Advanced)
Asystole V-Fib/V-Tach Couplet
Vent Rhythm PVC Bigeminy PVC Trigeminy
Tachy R on T PVC
Irr Rhythm Brady Missed Beat
Pacer not Pacing Vent Brady Pacer not Capture
VEB Run PVCs Acc. Vent Rhythm
IPVC Non-Sustain VT Multiform PVCs
Pauses/min High Pause Afib
PAC Bigeminy PVCs High Low Voltage(Limb)
ExtremeBrady PAC Trigeminy Wide QRS Tachy
Sustain VT ExtremeTachy V-Tach
Arrhythmia analyses
(Basic)
ASYSTOLE VFIB/VTAC COUPLET
VT > 2 BIGEMINY TRIGEMINY
VENT R on T PVC
TACHY BRADY MISSED BEATS
IRR VBRADY PNC
PNP
12-Lead ECG
Synchronization Analysis
Average parameters of heart beat
Heart rate (bpm)
Time limit of P wave (ms)
PR interval (ms)
QRS interval (ms)