Patient Monitor User Manual Product Specification
- 248 -
Range of Ventricular Rhythm
V-Tach
(Ventricular Tachycardia)
Advanced:
5 consecutive ventricular beats and ventricular HR 100 bpm.
Basic:
The interval of 5 consecutive ventricular beats is less than 600 ms
Vent Rhythm
(Ventricular Rhythm)
Advanced:
5 consecutive ventricular beats, and 20 bpm ventricular HR <
40 bpm.
Basic:
The interval of 5 consecutive ventricular beats ranges from 600 ms
to 1000 ms
Vent Brady
(VBRADY)
Advanced:
5 consecutive ventricular beats, and ventricular HR < 20 bpm.
Basic:
The interval of 5 consecutive ventricular beats is more than 1000 ms
Acc. Vent Rhythm 5 consecutive ventricular beats, and 40 bpm ventricular HR <
100 bpm.
Maximum Start-up Alarm Time for Tachycardia
V-Tach
(Ventricular Tachycardia)
1 mV 206 bpm
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
V-Tach
(Ventricular Tachycardia)
2 mV 195 bpm
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
Response Time of Heart
Rate Meter to Change in
HR
HR range: 80 bpm to 120 bpm
Range : Within 11 s
HR range: 80 bpm to 40 bpm
Range : Within 11 s
Tall T-wave Rejection
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17
minimum recommended 1.2 mV T-Wave amplitude