Patient Monitor User Manual Monitoring AG
- 173 -
WARNING
16 If, for whatever the reason, the IRMA probe is in direct contact with any parts of the
infant's body an insulation material shall be placed between the IRMA probe and the
body.
17 No modification of this equipment is allowed.
18 Measurements can be affected by mobile and RF communications equipment. It
should be assured that the IRMA probe is used in the electromagnetic environment
specified in this manual.
19 Disconnect the module from the airway adapter or set Work Mode to Standby when
the module is not in use. Setting path: AG Setup > Work Mode > Standby.
CAUTION
1 Never sterilize or immerse the IRMA probe in liquid.
2 The IRMA airway adapters are non-sterile devices. Do not autoclave the devices as
this will damage them.
3 Do not apply tension to the probe cable.
4 Do not operate the IRMA probe outside the specified operating temperature
environment.
5 (US Only) Caution: Federal law restricts this device to sale by or on the order of a
physician. For professional use. See instructions for use for full prescribing
information, including indications, contraindications, warnings, precautions and
adverse events.
NOTE:
For disposal of hospital waste such as accumulated fluids, calibration gases, sampled
gases, where not otherwise specified, follow local regulations regarding disposal of
hospital waste.
19.3 Monitoring Steps
19.3.1 Monitoring Steps for G7 Module
19.3.1.1 Zeroing the Sensor
EDAN G7 module itself has automatic zero function, only when the measurement is abnormal or
measurement results are doubtful, the user can perform manual zero as following steps:
1.
Wait until the monitor’s warm-up message disappears; keep the monitor away from AG
source.
2.
In the AG Setup menu, set Work Mode to Measure.
3.
Select Zero Calibration in AG Setup menu.