Patient Monitor User Manual Product Specification
- 257 -
Measure Parameter T1, T2, TD(the absolute value of T2 minus T1)
Channel 2
Sensor Type YSI-10K and YSI-2.252K
Unit
°C, °F
Measuring Range
0
°C to 50 °C(32 ºF to 122 ºF)
Resolution
0.1
°C (0.1 ºF)
☆Accuracy
1
±0.3 °C
Refresh Time Every 1 s to 2 s
Temperature Calibration At an interval of 5 to 10 minutes
Measuring Mode Direct Mode
Transient Response Time ≤ 30 s
Note 1: The accuracy consists of two parts, as following:
Accuracy (not including sensor): ± 0.1 °C
Sensor accuracy: ≤ ±0.2 °C
A.10 Quick TEMP
Only applicable to iM50 and M50.
Measuring Range
25
°C ~ 45 °C(monitoring mode)
35.5
°C~42 °C (prediction mode)
Operating Temp
10
°C ~ 40 °C
Sensor Type Oral/Axillary sensor, Rectal sensor
Resolution
0.1
°C
Accuracy
1
±0.3 °C
Response time < 60 s
Update time 1 s ~ 2 s
Warm-up time Less than 10 seconds
Prediction time Less than 30 seconds
Measuring Mode Direct Mode/Adjusted Mode
Transient Response Time ≤ 30 s
Clinical Bias
(-0.2 to -0.4 )
°C
Limits of Agreement 0.49