Patient Monitor User Manual Monitoring C.O.
- 164 -
Performing C.O. Measurement
1. Pick the C.O. Measure item in the C.O. Option menu. The C.O. Measure menu displays as
below:
C.O. Measure Window
1 Measurement curve 10 X axis: Change the Scale X (time) value. Two
modes are available: 0 s to 30 s, 0 s to 60 s. If
you start measurement in the 0 s to 30 s mode,
it will be switched to 0 s to 60 s mode
automatically if the measurement cannot finish
within 30 seconds. After the switch, no further
adjustment can be made to the Scale X.
2 Prompt message area
3 Cardiac Output
4 Cardiac Index
5 Body Surface Area
6 Blood Temperature 11 Y axis: Change the scale Y (temperature)
value. Three modes are available: 0 °C to 0.5
°C, 0 °C to 1 °C, 0 °C to 2.0 °C. Adjust the
scale by the temperature differences. A smaller
scale results in a larger curve.
7 Injectate Temperature
8 Start time of the measurement
9 Function keys
The functional keys on the C.O. measure window are explained in the following table:
Start
Start a measurement
Stop
If the blood temperature cannot resume in a considerably long time, the
measurement could not stop automatically. Use this button to stop the
measurement and display the C.O., CI calculation result.
Cancel
Cancel the processing measurement or cancel the result after measurement.
Record
Print out the curve.
Review
Enter the
Review window
1
3
4
5
6
7
8
11
9
2
10