Patient Monitor User Manual Warning-Score System
- 199 -
Value
3 2 1 0 1 2 3
RESP (rpm) 8 9~11 12~20 21~24 25
SpO
2
(%) 91 92~93 94~95 96
TEMP ()
35.0 35.1~36.0 36.1~38.0 38.1~39.0 39.1
SYS (mmHg) 90 91~100 101~110 111~219 220
HR (bpm) 40 41~50 51~90 91~110 111~130 131
Consciousness A V/P/U
Oxygen Yes No
In MEWS interface, select Criteria to check score criteria as following:
Value
3 2 1 0 1 2 3
HR (bpm) <40 41~50 51~100 101~110 111~129 130
SYS (mmhg) <70 71~80 81~100 101~199 200
RESP (rpm) <9 9~14 15~20 21~29 30
TEMP (°C)
<35.0 35.0~38.4 38.5
Consciousness A V P U
Note: (1) HR=40, value is 2; (2) SYS=70, value is 3.
The relationship between consciousness level and its display result is as below:
Consciousness Displayed Result
Sober A
Responsive to Voice V
Responsive to Pain P
Unresponsive U