EasyManua.ls Logo

EDAN iM70 - Page 258

EDAN iM70
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual Product Specification
- 246 -
Pace Pulse
Pulse Indicator
Pulse is marked if the requirements of IEC 60601-2-27: 2011,
Sect. 201.12.1.101.12 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to2.0 ms
Ascending time: 10 μs to 100 μs
Pulse Rejection
Pulse is rejected if the requirements of IEC 60601-2-27: 2011,
Sect. 201.12.1.101.13 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 μs to 100 μs
Pace Pulse Detecting Lead: one among I, II, III, AVR, AVL, AVF, V1, V2, V3,V4, V5, V6
Heart Rate
HR Calculation
Range
ADU: 15 bpm to 300 bpm
PED/NEO: 15 bpm to 350 bpm
Accuracy
±1% or 1 bpm, whichever is greater
Resolution 1 bpm
Sensitivity 300 μVPP
QRS Detection Range
The detection range has exceeded the requirement described in
the standard:
Width: 70 ms~120 ms for adult, 40 ms~120 ms for
Pediatric/neonate.
Amplitude: 0.5 mv~5 mv
In adult mode, these two signals are not responded:
1. when QRS amplitude of 0.15 mV or less is applied;
2. when QRS duration of 10 ms and QRS amplitude of 1 mV or
less is applied.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.15.
PVC
Range ADU: (0 to 300) PVCs/ min
PED/NEO: (0 to 350) PVCs/ min
Resolution 1 PVCs/min
Pause/min
Range ADU/PED/NEO: (0 to 30) pauses/min

Table of Contents

Related product manuals