Patient Monitor User Manual Product Specification
- 259 -
A.12 CO
2
Complies with ISO 80601-2-55: 2011.
EDAN G2 Sidestram Module
Intended
Patient
Adult, pediatric, neonatal
Measure
Parameters
EtCO
2
, FiCO
2
, AwRR
Unit mmHg, %, kPa
Measuring
Range
EtCO
2
0 mmHg to 150 mmHg (0 % to 20%)
FiCO
2
0 mmHg to 50 mmHg
AwRR 2 rpm to 150 rpm
Resolution
EtCO
2
1 mmHg
FiCO
2
1 mmHg
AwRR 1 rpm
Accuracy
EtCO
2
± 2 mmHg, 0 mmHg to 40
mmHg
Typical conditions:
Ambient temperature: (25
± 3) °C
Barometric pressure: (760
± 10) mmHg
Balance gas: N
2
Sample gas flowrate: 100 ml/min
± 5% of reading, 41 mmHg to
70 mmHg
± 8% of reading, 71 mmHg to
100 mmHg
± 10% of reading, 101 mmHg
to 150 mmHg
±12% of reading or
±
4 mmHg,
whichever is greater
All conditions
AwRR
± 1 rpm
Drift of
Measure
Accuracy
Meets the requirements of the measure accuracy
Sample Gas
Flowrate
70 ml/min or 100 ml/min (optional), accuracy:
±15 ml/min
Warm-upTime Display reading within 20 s; reach to the designed accuracy within 2 minutes.
Rise Time
< 400 ms (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 500 ms (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Response < 4 s (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)