EasyManua.ls Logo

EDAN iM70 - Page 279

EDAN iM70
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual Product Specification
- 267 -
- Adds less than 6 ml deadspace
- Pressure drop less than 0.3 cm H
2
O @ 30 LPM
Disposable infant:
- Adds less than 1 ml deadspace
- Pressure drop less than 1.3 cm H
2
O @ 10 LPM
(Infant Airway Adapter recommended for Tracheal Tube
ID size 4 mm)
Breath detection
Adaptive threshold, minimum 1 vol% change in CO
2
concentration.
Respiration rate
0 to 150
± 1 bpm. The respiration rate I displayed after
three breaths and the average value is updated every
breath. (Measured at I/E ratio 1:1 using breath simulator
according to EN ISO 80601-2-55 fig. 201.101.)
Fi and ET
FiCO
2
and EtCO
2
are displayed after one breath and have a
continuously updated breath average.
The following method is used to calculate end-tidal (ET)
values:
- The highest concentration of CO
2
during one breathing
cycle with a weight function applied to favor values closer
to the end of the cycle.
EtCO
2
will be within specification for all respiration rates
up to 150 bpm. (Measured at I/E ratio 1:1 using breath
simulator according to EN ISO 80601-2-55 fig. 201.101.)
Probe 2-9 channel NDIR type gas analyzer measuring at 4-10 μm.
Data acquisition rate 10 kHz (sample rate 20 Hz / channel).
Pressure, temperature and full spectral interference
correction.
Calibration No span calibration required for the IR bench.
Warm-up Time < 10 seconds (full accuracy)
Rise time (@ 10 l/min)
90 ms (measured @10 l/min with gas concentration steps
corresponding to 30% of total measuring range for each
gas)
Total system response time < 1 second (Measured according to EN ISO 80601-2-55)
AwRR
0 rpm to 150 rpm
AwRR Accuracy
± 1 rpm
Accuracy- Standard
Conditions
Gas Range Accuracy
CO
2
(0 to 15) vol% ±(0.2 vol% + 2% of reading)
The accuracy specifications above are valid for dry single gases at 22 ± 5 °C and 1013 ± 40
hPa.
Accuracy- All Conditions
CO
2
±(0.3 kPa + 4% of reading)
The accuracy specification above is valid for all specified environmental conditions except for
interference specified in Section Effects From Water Vapor Partial Pressure On Gas Readings

Table of Contents

Related product manuals