Patient Monitor User Manual Alarm Information
- 65 -
Message Cause Alarm Level Action Taken
Check
Watertrap/Sample
Line
(Dräger AG module)
Watertrap or sample line
falls off.
Low 1) Check whether
water trap is installed
normally.
2) Check whether
sample line is
installed normally.
AG Change
Watertrap
(Dräger AG module)
Malfunction in watertap Medium
Replace the
watertrap.
Watertrap will be
full
(Dräger AG module)
Watertrap will be full. Medium
Replace the
watertrap.
AG Agent Mixture
(Dräger AG module)
Mixture agents are detected,
but the monitor cannot
calculate MAC because of
low concentration.
Medium
Check agents’
concentration ratio.
CO
2
CO
2
Comm Fail
CO
2
module failure or
communication failure
High
Check if the water
tray has been fixed.
CO
2
Zero Required
Zero calibration failure Low
Disconnect the
sampling cannula or
adapter from the
airway; initiate the
zeroing before
making sure that no
expired air is inside
the sampling cannula
and adapter.