EasyManua.ls Logo

EDAN SONOTRAX Basic - Appendix 5 Overall Sensitivity; Sensitivity Measurement Data

EDAN SONOTRAX Basic
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SONOTRAX(2009) Series Ultrasonic Pocket Doppler User Manual
106
Appendix 5 Overall Sensitivity
Diameter of Target
Reflector
(mm)
Distance
(d)(mm)
A(dB)
Two-way Attenuation
B=∑B
a
+B
w
V
S
(r.m.s)
mV
V
n
(r.m.s)
mV
...
...
log20
10
smrV
smrV
C
n
s
dB
Overall
Sensitivity
(S=A(d)+B+C)
dB
∑B
(T: ultrasonic attenuation
phantom No. B
a
:dB)
B
w
(dB)
B
(dB)
1.58
A=45.7dB@2MHz
50
45.7
T
6#
6#
3#
-
0
57.5
72.15
34.32
6.45
109.6
B
a
24.9
24.9
7.7
-
75
45.7
T
6#
6#
2#
-
0
55.3
70.35
34.83
6.10
107.1
B
a
24.9
24.9
5.5
-
100
45.7
T
6#
6#
1#
-
0
53.5
72.62
35.64
6.30
105.5
B
a
24.9
24.9
3.7
-
200
45.7
T
6#
6#
-
-
0
49.8
75.47
35.86
6.24
101.7
B
a
24.9
24.9
-
-
2.38
A=43.2dB@2MHz
50
43.2
T
6#
6#
2#
1#
0
59.0
72.36
34.38
6.46
108.6
B
a
24.9
24.9
5.5
3.7
75
43.2
T
6#
6#
3#
-
0
57.5
74.31
34.83
6.58
107.2
B
a
24.9
24.9
7.7
-
100
43.2
T
6#
6#
2#
-
0
55.3
75.26
35.62
6.49
104.9
B
a
24.9
24.9
5.5
-
200
43.2
T
6#
6#
1#
-
0
53.5
75.42
35.83
6.45
103.2
B
a
24.9
24.9
3.7
-
Doppler Frequency (Hz)
333
Velocity of Target
(cm/s)
12.5
理邦保密文件
EDAN CONFIDENTIAL

Table of Contents