SONOTRAX(2009) Series Ultrasonic Pocket Doppler User Manual
62
5.3 Disinfection
In normal use the main unit does not need
disinfection. In case of being soiled, clean the main
unit case and then disinfect it by wiping it with a soft
cloth dampened with ethanol (75%) or isopropanol
(70%). Then wipe it dry with a dry cloth.
After each use, clean the probe and then disinfect it
by wiping it with a soft cloth dampened with ethanol
(75%) or isopropanol (70%). Then wipe it dry with a
dry cloth.
CAUTION
Pay attention not to immerse the probe socket
into the disinfector.
5.4 Sterilization
Do not sterilize the Doppler, unless this is necessary
according to your hospital regulation.
NOTE:
After cleaning or disinfection, check if the
Doppler function well. If any problem is detected,
please contact the manufacturer for service
before reusing it.