SONOTRAX(2009) Series Ultrasonic Pocket Doppler User Manual
79
Electrostatic
discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be
wood, concrete or
ceramic tile. If
floor are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Electrical Fast
Transient/Burst
IEC/EN61000-
4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
Not
applicable
Not applicable