Storage and Disposal
S900-01-880 Issue C
6 Storage and Disposal
6.1 Storage
CAUTION
Observe the storage temperature limits stated in Section 2.1. Storage below -30 oC (-22 oF) will permanently
damage the booster pump seals and oils.
Use the following procedure to store the booster pump:
1. Purge the vacuum system and the booster pump with dry nitrogen and disconnect the pump from the vacuum
system.
2. Shut down the pump as described in Section 4.4.
3. Disconnect the pump from the electrical supply and vacuum system.
4. Disassemble, clean and reassemble the pump. (Where necessary, contact your supplier or Edwards for advice.)
5. Place and secure protective covers on the inlet and outlet (exhaust) ports.
6. Store the pump in cool, dry conditions until it is required for use. When required, prepare and install the pump
as described in Section 3.
6.2 Disposal
Dispose of the booster pump and any components removed from it safely in accordance with all local and national
safety and environmental requirements.
Take particular care with components and waste oil which have been contaminated with dangerous process
substances.
Do not incinerate fluoroelastomer seals and 'O' rings.