E2M18 1.05 litres
* To operate the pump when the ambient temperature is outside the range specied in
Operang and storage condions on page 13 or to opmise pump performance when
processing condensible vapours, a dierent oil may be required.
3.5 Electrical data
Refer to Table: Electrical data: three-phase motors on page 17 and Table: Electrical data:
single-phase motors on page 17. For motor current informaon please refer to the
motor rang plate. The motor start-up current is drawn for less than one second, so
me-lag fuses must be used to prevent unnecessary fuse failure when the pump starts.
Fuses should be to EN60269 Performance on page 13. If using the pump at temperatures
lower than 13 °C, the start-up current will be drawn for longer; this may cause the motor
thermal overload device to open.
Table 5
Electrical data: three-phase motors
Pump item
number
Voltage (V)
Frequency
(Hz)
Full load
current
(A)
Recommended supply
protecon (A)
A34310940 200 - 220 50 3.3 10
380 - 415 50 1.9 6
200 - 230 60 2.3 10
460 60 1.6 6
A36310940 200 - 220 50 3.3 10
380 - 415 50 1.9 6
200 - 230 60 3.4 10
460 60 1.6 6
Motor output rang (connuous)
50 Hz operaon 0.65 kW
60 Hz operaon 0.75 kW
Table 6
Electrical data: single-phase motors
Pump item number
Voltage
(V)
Frequency
(Hz)
Full load current (A)
A36315903, A34316903,
A34315903
220 - 240
230 - 240
50
60
4.6
5.0
A34315981, A36315981 115
230
60
60
11.0
5.5
A34315904, A36315904 100
100 - 105
200
200 - 210
50
60
50
60
12.0
11.0
6.0
5.5
Page 17
A34310880_T - Technical data