1. Check that the oillevel in the oil sightglass is between the MAX and MIN level
marks on the bezel of the oil sightglass.
2. If the oillevel is near to or below the MIN level mark, remove the oil ller-plug and
pour more oil into the reservoir unl the oil reaches the MAX level mark. If the
oillevel goes above the MAX mark, remove the oil drainplug and drain the excess
oil from the pump. Ret the oil drainplug. Ret the oil llerplug.
3. If the oil is contaminated, drain and rell the pump with clean oil as described in
Replace the oil on page 38.
6.4 Replace the oil
Refer to Figure: General view of the pump on page 11 for the items in brackets.
1.
Operate the pump for approximately ten minutes to warm the oil, then switch o
the pump. (This lowers the viscosity of the oil and allows the oil to be drained from
the pump more easily).
2.
Isolate the pump from the electrical supply and disconnect it from the vacuum
system.
3.
Remove the oil ller-plug.
4.
Place a suitable block under the pump-motor to lt the pump and place a suitable
container under the oil drain-plug (gravity drain). Remove the oil drain-plug and
allow the oil to drain into the container.
5. If the oil is dirty or contaminated:
Ret the oil drain-plug and pour clean oil into the pump.
Reconnect the pump to the electrical supply and operate the pump for about
5 to 10 minutes.
Disconnect the pump from the electrical supply, remove the oil drain-plug and
allow the oil to drain out of the pump.
Repeat this step unl the oil reservoir is clean.
6. Ret the oil drain-plug, remove the block and reconnect the pump to the vacuum
system.
7. Fill a suitable container with clean oil and pour the oil into the ller hole unl the
oil-level reaches the MAX level mark on the bezel of the oil sight-glass.
8.
Allow a few minutes for the oil to drain into the pump. If necessary, add more oil.
Ret the oil ller-plug.
6.5 Inspect and clean the inlet-lter
Refer to Figure: Remove/replace the inlet lter on page 39.
1.
Unscrew the inlet adaptor and remove the O-ring, circlip and inlet-lter.
2.
Wash the lter in a suitable cleaning soluon. Allow the lter to dry.
3.
Ret the inlet-lter, circlip, O-ring and inlet adaptor.
Page 38
A34310880_T - Maintenance

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards E2M18 and is the answer not in the manual?

Edwards E2M18 Specifications

General IconGeneral
BrandEdwards
ModelE2M18
CategoryWater Pump
LanguageEnglish

Related product manuals