Technical Data
A533-23-880 Issue C
2.6 Electrical data
Electrical supply
Supply voltage 200-208 V at 50/60 Hz, 3-phase
440 V at 60 Hz, 3-phase
Voltage tolerance 10% (except for 208 V at
50 Hz, which are +6%)
Installation Category Category 2 (EN61010)
Pollution Category Category 1 (EN61010)
Full load current ratings refer to Table 5
Fuse types and ratings
F1
*
*
These fuse rating are included for information only. You cannot change these fuses; a
Edwards service engineer must change these fuses.
30 A, 600 V, Class J
F2
*
30 A, 600 V, Class J
F3
*
30 A, 600 V, Class J
F4
*
20 A, 600 V, Class J
F5
*
20 A, 600 V, Class J
F6
*
20 A, 600 V, Class J
F7 1.6 A, 250 V
F8 1.6 A, 250 V
F9 3.15 A, 250 V
Emergency stop switch rating 24 V a.c./d.c., 5 A max
External emergency stop switch
Rating 24 V a.c., 5 A
Load 700 mA (inductive)
Table 5 - Full load current ratings
Supply voltage and frequency
200-208 V
50 Hz
200-208 V
60 Hz
440V
60Hz
FDP
Full load (A)
Motor rating (kW)
22.1
5.5
21.7
5.5
10.1
5.5
HMB
Full load (A)
Motor rating (kW)
14.5
2.9
15.4
3.5
7.2
3.5
Maximum input power to iF system (kW) 11.4 12.1 12.1