To connect the electrical supply to the system through the connector supplied with the
pump, use the procedur
e as follows. Refer to: Figure: Mains input supply connector -
Han®40A (iXM200, iXM600 and iXM1200) to Figure: Mains input supply connector -
Han® K 4/4 (iXM3000) for your connector type. To use a dierent connector than that
supplied, follow the manufacturer's instrucons.
When you prepare the electrical supply cable, make sure that the earth (ground)
conductor is longer than the phase conductors. This will make sure that if the cable is
accidentally dragged and the strain relief bush on the electrical supply connector mang
half fails, the earth (ground) conductor will be the last conductor to be pulled from the
connector.
1. Aach the cable gland onto the connector hood.
2. Put a applicable cable, through the cable gland and hood. The cross seconal area
of the cable wires should be 6 mm
2
. Cables must be bare ended without ferrules
for correct clamping in the connector block. Refer to Table: Wire assembly
according to VDE 0295
3. Aach the coding pins to the connector block according to the branch supply
v
oltage as shown in Figure: Coding pin conguraon for connector type (200 V
supply for low volts, 400 V supply for high volts).
4. Before you start the assembly, use the hex (Allen) key as shown in Figure: Method
f
or connecng phase wires or Figure: Mains input supply connector - Han® K 4/4
(iXM3000). Make sure that the axial cone is screwed fully anclockwise to
completely open the contact chamber.
5. For the three live conductors (L1, L2, L3), carefully remove the cable insulaon to
the exact dimension specied in Figure: Method for connecng phase wires. Do
not twist the cable strands.
6. Refer to Figure: Mains input supply connector - Han®40A (iXM200, iXM600 and
iXM1200) to idenfy the connecons. Connect the three live conductors to the
connector block: L1 to a1, L2 to a2 and L3 to b1. Insert each wire completely into
the contact chamber unl the copper stands reach the boom.
7. Tighten the connecon using an appr
opriate Allen key as shown to a torque of
2 Nm. Hold the cable in posion while applying the recommended ghtening
torque. Refer to Figure: Mains input supply connector - Han®40A (iXM200, iXM600
and iXM1200) and Figure: Mains input supply connector - Han® K 4/4 (iXM3000).
8. Aach the earth (ground) wire to the protecve earth connecon on the
connector block as shown in Figure: Mains input supply connector - Han®40A
(iXM200, iXM600 and iXM1200) using the following:
Prepare the end of the 6 mm
2
earth wire t
o a strip length of 10 mm and twist
the strands.
Insert the wire into the earth terminal block.
Tighten the connecon using a at blade screwdriver.
9. Install the outer cover to the connector block then ghten the cable gland
10. Connect the mang half to the electric
al supply connector of the dry pump system
(Figure: Controls/connectors on the rear of the system).
11. Aach a secondary protecve earth (ground) conductor (with a cross-seconal
area at least equal to phase conductor size) to the protecve earth (ground) stud.
Refer to Figure: Mains connector on pump bulkhead.
10/2020 - ©Edwards Limited
Page 39M56635880_D
M56635880_D - Inst
allaon

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards iXM Series and is the answer not in the manual?

Edwards iXM Series Specifications

General IconGeneral
BrandEdwards
ModeliXM Series
CategoryWater Pump
LanguageEnglish

Related product manuals