11 Technical Data
Page
48
/
52
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various
healthcare locations where medical equipment and/or systems are
used. When they are used in close proximity to medical equipment
and/or systems, the medical equipment and/or systems’ basic
safety and essential performance may be affected. The E-xtreme has
been tested with the immunity test level in the below table and meet
the related requirements of IEC 60601-1-2:2014. The customer and/or
user should help keep a minimum distance between RF wireless
communications equipments and the E-xtreme as recommended
below.
Test
frequ-
ency
(MHz)
Band
(MH-
z)
Service
Modu-
lation
Maxim-
um
power
(W)
Dist-
ance
(m)
Immu-
nity
test
level
(V/m)
385
380-
390
TETRA
400
Pulse
modu-
lation
18Hz
1.8
0.3
27
450
430-
470
GMRS
460
FRS 460
FM
± 5
kHz
deviat-
ion
1 kHz
sine
2
0.3
28
710
704-
787
LTE
Band 13,
17
Pulse
modu-
lation
217Hz
0.2
0.3
9
745
780
810
800-
960
GSM
800/900
,
TETRA
800,
iDEN
820,
Pulse
modu-
lation
18Hz
2
0.3
28
870