3 Before Use
Page 6 / 28
Before Use
Scope of application
Ultra X is used in endodontic treatment by application of ultrasonic energy. The
Ultra X can provide the energy for tip oscillation and vibration in frequency
(45KHz+5KHz) required to create sufficient acoustical streaming and cavitation
necessary to effectively clean, penetrate, and remove vapor lock. A cleaned root
canal system makes for better outcomes and reduces retreatment rate.
This device must only be used in hospital environments, clinics or dental offices
by qualified dental personnel and not used in the oxygen-rich environment.
Contraindications
The Ultra X is contraindicated in cases where patient/user carry medical
implants such as pace makers or cochlear implants etc.
Do not use the device for implants or other non-endodontic dental procedures.
Safety and effectiveness have not been established in pregnant women and
children.
Read the following warnings before use:
1. The device must not be placed in humid surroundings or anywhere where it can
come into contact with any type of liquids.
2. Do not expose the device to direct or indirect heat sources. The device must be
operated and stored in a safe environment.
3. The device requires special precautions with regard to electromagnetic
compatibility (EMC) and must be installed and operated in strict compliance with the
EMC information. In particular, do not use the device in the vicinity of fluorescent
lamps, radio transmitters, remote controls and do not use this system near the active
HF Surgical Equipment in the hospital. Portable RF communications equipment
(including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the E-connect S, including