The instructions provided above have been validated by the manufacturer of
the medical device as being capable of preparing a medical device for use.
It remains the responsibility of the processor to ensure that the processing,
as actually performed using equipment, materials and personnel in the
processing facility, achieves the desired result. This requires verification
and/or validation and routine monitoring of the process. Likewise, any
deviation by the processor from the instructions provided should be properly
evaluated for effectiveness and potential adverse consequences.
Disinfection components
Handpiece
Handpiece Base
Adapter
Wipe all the surfaces with a cloth lightly moistened with Ethanol for Disinfection
(Ethanol 70 to 80vol%) at least 2min, repeat for 5 times.
NOTE
Do not use anything except Ethanol for Disinfection (Ethanol 70 to 80 vol%).
Do not use too much ethanol as it’s going into machine and damage the
components inside.