EMC Tables
Guidance and manufacturer’s declaration – electromagnetic emissions
The Ultra X is intended for use in the electromagnetic environment specified below.
The customer or the user of the Ultra X should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Ultra X uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Ultra X is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Ultra X is intended for use in the electromagnetic environment specified
below. The customer or the user of the Ultra X should assure that it is used in
such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
+/- 8 kV contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transients/bursts
IEC 61000-4-4
±2kV
100kHz repetition
frequency
±2kV
100kHz repetition
frequency
Mains power quality should
be that of a typical
commercial or hospital
environment.