EMC Information
The performance of the EX3242-3D ensures appropriate display of images.
Intended Use Environments
The EX3242-3D is intended to be used in professional healthcare facility environments such as clinics
and hospitals (including use in the vicinity of high-frequency surgical equipment such as
electrosurgical knives).
The following environments are not suitable for using the EX3242-3D.
Home healthcare environments
In the vicinity of short-wave therapy equipment
RF shielded room of MRI medical equipment systems
In shielded special environments
Installed in vehicles including ambulances
Other special environments
WARNING
The EX3242-3D requires special precautions regarding EMC and during installation. You need to
carefully read the EMC Information and the “PRECAUTIONS” section of this document, and observe
the following instructions when installing and operating the product.
The EX3242-3D should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to check for normal
operation in the conguration in which it will be used.
When using portable RF communication equipment, keep it 30 cm (12 inches) or more away from
any part, including cables, of the EX3242-3D. Otherwise, degradation of the performance of this
equipment could result.
Anyone who connects additional equipment to the signal input parts or signal output parts when
conguring a medical system is responsible for ensuring that the system complies with the
requirements of IEC/EN60601-1-2.
Images may be distorted if the product is used near a device such as high-frequency surgical
equipment. Check in advance to ensure that no problems occur during use.
Be sure to use cables that satisfy the following requirements.
Use of cables that do not satisfy the requirements could result in increased electromagnetic
emissions, reduced electromagnetic immunity of this equipment, and incorrect operation.
Cables Max. Cable Length Shielding
AC Cord 2 m Unshielded
DC Cord 17.5 m Shielded
BNC Cable (SDI) 30 m Shielded
DVI Cable 5 m Shielded
DisplayPort Cable 5 m Shielded
HDMI Cable 5 m Shielded
RS-232C Cable 5 m Shielded
USB Cable 5 m Shielded
27
Appendix

Other manuals for Eizo CuratOR EX3242-3D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eizo CuratOR EX3242-3D and is the answer not in the manual?

Eizo CuratOR EX3242-3D Specifications

General IconGeneral
BrandEizo
ModelCuratOR EX3242-3D
CategoryMonitor
LanguageEnglish

Related product manuals