PRODUCT MAINTENANCE
06/2020 45 NP-DK50 2V 50, 2x2V 110-A-5_06-2020-MD
PRODUCT MAINTENANCE
17. PRODUCT MAINTENANCE
The operator should carry out device checks regularly in the intervals defined by
applicable regulations. Test results must be recorded.
The device has been designed and manufactured to keep its maintenance to a minimum. For correct
and reliable operation of the compressor perform the following operations:
Risk of servicing by persons without required qualification.
Repair works beyond standard maintenance (see chapter.17.1) may only be
performed by a qualified technician (organization authorized by the
manufacturer) or manufacturer’s customer service.
Standard maintenance works (see chapter 17.1) may only be carried out by
a trained operator.
Only use spare parts and accessories approved by the manufacturer.
Risk of injury and damage to the device.
Before any maintenance work, it is necessary to:
- check, if it is possible to disconnect compressor from the supplied
equipment, so that there is no risk of damage to a person using the given
appliance or any other material damages
- switch off the compressor
- disconnect it from the mains (pull out the mains plug)
- vent the compressed air from the air tank
Risk of injury when venting the compressed air.
Protect eyesight – wear goggles when venting the compressed air from the
compressor pneumatic system (air tank).
Pump components (head, cylinder, pressure hose) have high temperature during
and shortly after compressor operation – do not touch these components!
Let the device cool before any product maintenance, service or connection/
disconnection of pressurized air!
The removed grounding conductor during service must be connected back to the
original position after completing the service.
For any maintenance or repair work, the compressor may be removed from the cabinet by
pulling it from the cabinet (on rolling casters as much as the cord between the compressor
and the cabinet allows). See also chap. 17). The required maintenance or repair work may
then be performed.
If the cord between the compressor and the cabinet becomes disconnected (by pulling the
mains plug, chap. 12.2), the compressor shall not be functional. In order to bring the
compressor into function again, please follow chap. 12.2 or 17.10 or 0.