ļ›™ 2002 by ELTRA GmbH Germany – January 2002 – Service Manual OH 900 / ON 900 / ONH 2000
4.3-39
GAS flow diagrams ONH-2000
from S/N 1736xxxxxx
C O
2
H O
2
3 0l /h
impulse
Flow sensor
Infrared Cell
gas stop
Carrier
Pressure regulator
2
N
25186
Purge
valve
He
1 5 l / h
01160
Pressure regulator 0,35bar/5,3psi
5 0 l /h
Ele ctrode
fu rnace
Controlled
1. 5
b a r
0. 5
0 2. 5
2
1
11430
ON-Operation
T C C e l l
green blue
black
red
Reference flow
V3
Reference flow outlet
11435
OH-Operation
2-4bar
24VDC
22 ps i
2 2ps i
yellow
grey
ON-Operation
Measur e
ONH-2000
D us t
H O
2
C O
2
C O
2
H O
2
N
ON-Operation
Analysis mode
OUT
Gas pump
IN
Carrier gas outlet
Energised solenoid valve coils.
30l/h
50l/h
Pressure
switch
H
11-7
V1
V4
25195
V8
yellow
PR4
V6
V2
230VAC
V5
1 ,5ba r
30 -60psi
3 0-60psi
2-4bar
1,5b ar
22 psi
1,5 bar
PR2
24V DC
OH-Operation
01160
24V DC
CO CO furnace
2
25195
V7
V2
V3
V5
V6
V1
V4
V8
V7
IN
IN
OUT
OUT

Table of Contents