2002 by ELTRA GmbH Germany – January 2002 – Service Manual OH 900 / ON 900 / ONH 2000
2.8-1
2.8
2.8
Furnace lift adjustment
The purpose of this adjustment is to
ensure that the furnace can lift and
fit correctly into the upper chamber.
To adjust, open the furnace
and loosen the screws ( C ),
the furnace is now free to move
horizontally.
Close the furnace, the furnace aligns
itself as it moves into the upper chamber.
Fasten the screws ( C ) again.
B
A
C
2-8-1