2002 by ELTRA GmbH Germany – January 2002 – Service Manual OH 900 / ON 900 / ONH 2000
2.14-1
2.14
2.14
Temperature regulation of the TC Cell:
The thermostabilisation of the TC cell is similar to the IR cell.
Adjust the set point of the thermostabilisation with trimmer No.1.
Turn clockwise to increase the temperature.
Caution :
The influence of the temperature on the baseline is very high. Wait for the stabilisation
after every adjustment.
TC CELL
3
1
2
2-12
04000