EasyManua.ls Logo

EM TEST CWS 500N2 - Calibration Setup EM TEST and IEC 61000-4-6

EM TEST CWS 500N2
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EM Test CWS500 N2
User manual V 3.00 22 / 38
5.5.2.
Calibration Setup EM TEST and IEC 61000-4-6
The setup for calibration according to IEC 61000-4-6 is shown in the figure 5.7. below. EM Test has carefully
tested the calibration setup and propose to eliminate the 150
terminating resistor on the AE port for CDN
calibration. ( Fig 5.8)
Fig 5.7 : Calibration Setup per IEC 61000-4-6
Fig 5.8 : Calibration Setup EM Test method
Reason for the EM Test calibration test setup:
- The AE port must only be applied to unshielded cables (shielded cables have their shielding connected to the
ground reference plane of the AE port).
- The uncoupling inductance according to the standard is >280
H. The EM Test CDN has a higher inductance
(CDN M3 typical >500
H). Therefore, the calibration error at 150kHz is max 10% and will be reduced by
increasing the frequency. (Error at 400kHz approx.5%).
- The tolerance of the applied signal according to IEC 61000-4-6 is
25%. The figure below shows two CDN M3
10V measurements with different calibration methods. The results are within the
25% limit and the result is
very similar.
- The influence of the internal synthesizer attenuator is responsible for the non-linear result of the calibration
test.
EM Test Calibration Setup
EUT port : 1x 150:
R100 ( 100 )
power meter ( 50 )
AE port : open circuit
IEC 61000-4-6 Setup
EUT port : 1x 150:
R100 ( 100 )
power meter ( 50 )
AE port : 1x 150:
R100 ( 100 )
terminating ( 50 )
Fig 5.9 : Difference between the calibration methods
However, for a calibration according to the standard it is necessary to add the additional calibration acessories
for the AE port termination :
150 to 100 adapter : R-100
50 termination resistor : T-50

Table of Contents

Related product manuals