EM Test CWS500 N2
User manual V 3.00 32 / 38
CWS500N2.1
Summary:
Output power 150W (nominal)
Output impedance 50 Ohm, VSWR 2.0: 1 nominal
Level approx. -13dBm ... 51.7dBm
Current 4MHz ... 400MHz max. 950mA (150W amplifier, without attenuator, F-130A clamp)
Test frequencies ( synthesizer )
Sinusoidal (CW) 9kHz - 1000MHz
Frequency bands 9kHz - 1000MHz step: 1kHz ( icd.control and Quickstart online)
Quickstart setting 9kHz - 9,999MHz step: 1kHz
10MHz - 99,99MHz step: 9kHz
100MHz - 1000MHz step: 100kHz
Modulation
AM Frequency: 1Hz ... 3000Hz
Index: 0%...95%
PM Frequency: 1Hz ... 3000Hz
Duty cycle: 10%...80%
Attenuator
Range 0 dB - -63.5dB
Step level 0.5dB
Timing:
Dwell time general td = 0,3s - 9999.9s
pulse modulation td = 3s - 9999.9s
Rest time tr = 0 / 0,3s - 9999.9s
Output
Direct N type connector
EUT control
BNC input FAIL 1 Fail 1; test will be stopped immediately
BNC input FAIL 2 Fail 2; failure is detected, test continues
Routines (integrated)
Service Service, Setup
Test routines All test only in remote control with icd.control software version >3.00