A.2 K-Factor and inverse K-Factor
Equation A-1: Frequency volumetric flow rate K-Factor
KFactor =
FreqQ
FullScale
(MaxFreq)3600sir
*
and
Equation A-2: Frequency volumetric flow rate inverse K-Factor
InvKFactor =
(MaxFreq)(3600sir)
FreqQ
FullScale
*
where
KFactor = frequency “K-Factor” (pulses/volume**) (Freq1KFactor and Freq2KFactor)
InvKFactor = frequency “Inverse K-Factor” (volume**/pulse) (Freq1InvKFactor and
Freq2InvKFactor)
FreqQ
FullScale
= frequency full-scale volumetric flow rate (volume**/time unit*)
(Freq1FullScaleVolFlowRate and Freq2FullScaleVolFlowRate)
MaxFreq = maximum frequency (Hz = pulses/time unit*) (Freq1MaxFrequency and
Freq2MaxFrequency)
* TimeUnit = time conversion factor depends on the VolFlowRate Time Unit data point:
volume/second = 1 s/s
volume/minute = 60 s/m
volume/hour = 3600 s/h
volume/day = 86400 s/d
** Volume = where the volume is selected via data points:
Units System
VolUnitUS
gallons
barrels
VolUnitMetric
cubic meters
liters
Conversion factors Maintenance and Troubleshooting manual
June 2019 P/N 3-9000-769
104 Gas Ultrasonic Flow Meters

Other manuals for Emerson Daniel 3410 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Daniel 3410 Series and is the answer not in the manual?

Emerson Daniel 3410 Series Specifications

General IconGeneral
BrandEmerson
ModelDaniel 3410 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals