3.7 T-200 transducer housing removal and
installation
WARNING
CONTENTS MAY BE HAZARDOUS
The meter must be fully depressurized and drained before attempting to remove the
T-200 transducer assembly. If gas or fluid begins to leak, stop immediately and reinstall
the transducer assembly. Failure to comply may cause serious injury or equipment
damage.
Figure 3-18: Components of T-200 transducer assembly
This section provides the procedures for removing a transducer housing from a T-200
transducer assembly and installing a transducer housing on a transducer stalk. If the
transducer assembly is installed in a meter body, follow the procedure in Section 3.5.1 to
remove the assembly from the meter body first. After installing or replacing a transducer
housing, follow Section 3.5.2 to install the transducer assembly to the meter body.
3.7.1
T-200 transducer housing removal
This procedure is for removing a transducer housing from a T-200 transducer assembly
installed in a meter body.
WARNING
CUTTING HAZARD
Sharp edges may be present on the transducer retaining ring. Wear appropriate eye
protection equipment when removing or installing the transducer retaining ring. Failure
to comply may cause serious injury.
Procedure
1. Follow Section 3.5.1 to remove the transducer assembly from the meter body first.
Ensure that the line pressure is down to atmospheric pressure prior to disassembly.
2. Hold the T-200 transducer assembly vertically with the retainer facing up.
3. Remove the transducer retainer from the stalk. Do not drop the transducer capsule.
4. Carefully pull the transducer capsule out of the stalk and place it in a safe place.
Maintenance and Troubleshooting manual Meter repairs
P/N 3-9000-769 June 2019
Maintenance and Troubleshooting manual 79