EasyManua.ls Logo

Emerson MICRO MOTION 2700 - 9.2.2 View process variables using ProLink III

Emerson MICRO MOTION 2700
374 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.3 Set up the Weights & Measures application
using ProLink III
When your transmitter is set up to comply with OIML or NTEP requirements, and the
transmitter is “secured,” selected measurement data from the transmitter is approved for
Weights & Measures applications.
Prerequisites
Important
Review the Weights & Measures requirements for your location and ensure that you are complying
with all local requirements as well as the OIML or NTEP requirements.
If necessary for your installation, arrange for a site visit by a certified Weights & Measures
inspector, and ensure that the inspector is present for the appropriate portions of this
procedure.
Ensure that the transmitter is ready to be secured, i.e., it is configured as desired and you
have performed all appropriate tests and adjustments. After the transmitter is secured,
you cannot make any changes to configuration, many maintenance actions are disallowed,
and some operator actions are disallowed.
Procedure
1. Make a service port connection from ProLink III to your transmitter.
2. Choose Device Tools > Configuration > Weights & Measures and set Regulatory Agency to the
apropriate regulatory agency for your application.
Option Description
NTEP
Regulatory agency for the U.S.A. and Canada
OIML
Regulatory agency for all other world areas
3. Choose Device Tools > Configuration > Totalizer Control Methods and set totalizer control
methods as desired.
Parameter Option
Reset Totalizers from Display
Enabled: You can reset totalizers from the display,
whether or not the transmitter is secured.
Disabled: You can reset totalizers from the display only
when the transmitter is unsecured.
Reset Totalizers via Remote
Communications
Enabled: You can reset totalizers using digital communi-
cations, whether or not the transmitter is secured.
Disabled: You can reset totalizers using digital communi-
cations only when the transmitter is unsecured.
Set up the Weights & Measures application
Configuration and Use Manual 143

Table of Contents

Other manuals for Emerson MICRO MOTION 2700

Related product manuals