EasyManua.ls Logo

Emerson MICRO MOTION 2700 - Page 76

Emerson MICRO MOTION 2700
374 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Derived variables and calculated process variables Table 4-15:
Derived Variable
Description
Calculated process variables
Density at
reference
tempera-
ture
Standard
volume
flow rate
Specific
gravity
Concen-
tration
Net mass
flow rate
Net vol-
ume flow
rate
Density at reference
temperature
Mass/unit volume, cor-
rected to a given refer-
ence temperature
Specific gravity
The ratio of the density
of a process fluid at a
given temperature to
the density of water at
a given temperature.
The two given temper-
ature conditions do not
need to be the same.
Mass concentration
derived from refer-
ence density
The percent mass of
solute or of material in
suspension in the total
solution, derived from
reference density
Mass concentration
derived from specif-
ic gravity
The percent mass of
solute or of material in
suspension in the total
solution, derived from
specific gravity
Volume concentra-
tion derived from
reference density
The percent volume of
solute or of material in
suspension in the total
solution, derived from
reference density
Volume concentra-
tion derived from
specific gravity
The percent volume of
solute or of material in
suspension in the total
solution, derived from
specific gravity
Concentration de-
rived from reference
density
The mass, volume,
weight, or number of
moles of solute or of
material in suspension
in proportion to the to-
tal solution, derived
from reference density
Configure process measurement
68 Micro Motion
®
Model 2700 Transmitters with Analog Outputs

Table of Contents

Other manuals for Emerson MICRO MOTION 2700

Related product manuals